| Wavelengths |
254,
280, 365, 420, 505 |
| Lamps |
Low
Pressure: Hg 254 and 280; separate lamps for 365, 420, 505 |
| Spectral
Resolution |
Dependent
on Filter Selection |
| Linearity |
Better
than 2% |
| Wavelength
Accuracy |
N/A |
| Wavelength
Reproducibility |
N/A |
| Drift |
<5.0
x 10(-4) AU/Hour |
| Noise |
1.0
x 10(-5) AU@254nm |
| Recorder
Ranges |
0.005,
0.01, 0.02, 0.05, 0.10, 0.20, 0.50 and 1.0 AU Full Scale |
| Recorder
Output |
10mV |
| Integrator/Data Output |
1.0 V/AU |
| Display |
None |
| Dimensions |
11.3"L x 7.5"W x 3.0"H |
| Weight |
6lb (2.7 Kg) |
| Power |
115, 230 VAC, 50/60 Hz |
|
| Flowcells |
| Analytical |
7.0mm/10µL
Volume; PEEK or SS Body, Quartz window |
| Semi-Prep |
2.5mm/4.4µL
Volume |
| Preparative |
2.0mm/4.0µL
Volume, PEEK |
| Microbore |
5.0mm/1.5
µL Volume, PEEK |
|